Mass Spectrometry and Peptide Identity Testing Explained
Kineris Labs · Updated 27 October 2026 · 2 min read · 3 references
A purity figure tells you that one component dominates a sample. Mass spectrometry tells you whether that component is the molecule on the label. This guide explains how identity testing works, how to read the result on a Certificate of Analysis, and what it can miss.
This article summarises published scientific literature for research context. It is not medical advice and makes no claim about the effects of any Kineris product. Findings described here come from the cited studies, most of them in animal or cell models. All Kineris products are supplied for laboratory research use only and are not for human or veterinary use.
How it works
A mass spectrometer turns molecules into gas-phase ions, separates them by mass-to-charge ratio (m/z) and counts them. For peptides the usual ionisation methods are electrospray ionisation (ESI) and matrix-assisted laser desorption/ionisation (MALDI), the two techniques that made mass spectrometry of large biomolecules routine.[1] Coupled to liquid chromatography (LC-MS), it can measure the mass of the main HPLC peak and of the impurities beside it.
Reading the result
The theoretical mass is calculated from the peptide's molecular formula. The CoA should report the measured value next to it. Electrospray adds protons, so a peptide of mass M appears at (M + nH)/n for several values of n. BPC-157, with an average molecular weight of about 1419.5 g/mol, would typically show a doubly charged ion near m/z 711 and a triply charged ion near m/z 474. Deconvolution software combines these into one measured mass, which should agree with the theoretical value within the instrument's stated tolerance.
| Result | Interpretation |
|---|---|
| Measured mass matches theoretical mass | Consistent with the expected sequence |
| Difference of about −18 Da or −17 Da | Loss of water or ammonia, often a minor related impurity or an artefact |
| Difference of +16 Da | Oxidation, commonly of methionine or tryptophan |
| Difference of one residue mass (for example −57 Da for glycine) | Deletion sequence |
| Mass far from the expected value | A different compound or fragment |
These mass shifts correspond to the impurity classes that peptide synthesis produces, which is why mass spectrometry is central to profiling impurities in synthetic peptides.[2]
Why it matters in practice
Identity testing is how wrong products are found. When Esposito and colleagues analysed a product sold as "TB-500", mass spectrometry showed it was a seven-residue acetylated fragment of thymosin beta-4 (Ac-LKKTETQ), not the 43-residue protein.[3] An HPLC trace of that product could show high purity, because the sample was mostly one molecule. It was just not the molecule many buyers expected.
Limits of an identity test
- Isomers have the same mass. A sequence with two residues swapped, or a racemised residue, can pass a simple mass check.[2]
- A mass check is not sequencing. Tandem MS (MS/MS) fragments the peptide to confirm the order of residues and is more definitive.[1]
- MS is not quantitative on its own. It confirms what is present, not how much.
Together with HPLC purity, an identity result is the core of a useful Certificate of Analysis.